== Exhaustive Photon Restoration ==
 
== Exhaustive Photon Restoration ==
 +
 +
Exhaustive Photon Restoration (EPR) is BIG's 3D Deconvolution routine. It uses a point spread function (PSF) to deconvolve images from a microscope. The PSF is a 3D image of a point source, usually a small fluorescent bead. The PSF should be acquired under similar conditions as the images that are trying to be deconvolved. The fourier transform of the PSF represents the optical transfer function (OTF) of a microscope and is used to deconvolve images.
  
 
There are three main programs involved in 3D Deconvolution: '''preppsf''', '''prepdata''', and '''epr_i2i'''. They all use our own i2i image format.  
 
There are three main programs involved in 3D Deconvolution: '''preppsf''', '''prepdata''', and '''epr_i2i'''. They all use our own i2i image format.  

Revision as of 19:00, 20 November 2018


Exhaustive Photon Restoration

Exhaustive Photon Restoration (EPR) is BIG's 3D Deconvolution routine. It uses a point spread function (PSF) to deconvolve images from a microscope. The PSF is a 3D image of a point source, usually a small fluorescent bead. The PSF should be acquired under similar conditions as the images that are trying to be deconvolved. The fourier transform of the PSF represents the optical transfer function (OTF) of a microscope and is used to deconvolve images.

There are three main programs involved in 3D Deconvolution: preppsf, prepdata, and epr_i2i. They all use our own i2i image format.

preppsf prepares the psf for epr_i2i.

prepdata prepares the images for epr_i2i.

epr_i2i performs the 3D deconvolution.

preppsf

Run the command for full option list: preppsf -h

NAME

   preppsf - prepares a 3-D point spread function (psf) image data set, 
   acquired using a Digital Imaging (light) Microscope, for use with image
   restoration/3-D reconstruction. Normally psf images are first processed
   using prepdata (see below) to apply basic corrections and any background
   corrections, but not temporal corrections.

SYNOPSIS

   preppsf [options] before-image after-image

DESCRIPTION

   preppsf performs further corrections on a psf image data set based on
   the fluorescence D.I.M image formation/acquisition model. The before-
   image argument is the name of a 3-D psf image set, acquired (DIM-1,
   DIM-2, CELLscan, UFM, or other)  and processed using prepdata (see below)
   without the -norm option. The psf is extracted as a symmetric sub-region
   (square in XY) centered at the psf origin (-center), optionally
   normalized for constant total intensity (-norm), and optionally masked
   (-mask) to exclude any extramural data. The after-image argument is the
   file name for the processed psf image set. All options must precede
   input/output image files

Important Information

  • Numerical Aperture of the objective.
  • Index of refraction of immersion medium.
  • Effective Pixel Size at the object in micrometers/pixel
  • DIstance Spacing among Z planes in micrometers.


Example Usage

preppsf -NA 1.4 -n 1.515 -size 0.187 -spaced 0.25 -norm infocus -mask 1 psf_in  psf_out

prepdata

Run the command for full option list: prepdata -h

NAME

   prepdata - prepares a 2-D, 3-D or 4-D image data sets, acquired using a                                                                                                      
   Digital Imaging (light) Microscope, for further processing (e.g. image                                                                                                       
   restoration/3-D reconstruction)                                                                                                                                                 
                                                                                                                                                                                   

SYNOPSIS

   prepdata [options] before-image after-image                                                                                                                                     
                                                                                                                                                                                      

DESCRIPTION

   prepdata performs all necessary corrections on a image data set based on
   the fluorescence D.I.M image formation/acquisition model, and can be
   used to perform basic imaging corrections to many forms of digitally
   acquired image data. The before-image argument is the name of a 3-D image
   set as acquired (DIM-1, DIM-2, CELLscan, UFM, or other). The after-image
   argument is the file name for the corrected image set, ready for further
   processing. The normal order of application of corrections (see -before)
   is basic[->background[->temporal]]. All calculations are performed using
   single-precision floating-point arithmetic. All options must precede
   input and output image file names.

epr_i2i

Run the command for full option list: epr_i2i -h

NAME

   epr_i2i - Exhaustive Photon Replacement (EPR) restores contrast by removing 
   residual out-of-focus light and improves resolution while maintaining 
   numerical accuracy of 3-D images of specimens obtained with serial 
   optical sectioning from wide-field or confocal light microscopy.

SYNOPSIS

   epr_i2i [options] before-image[.i2i] after-image[.i2i]

DESCRIPTION

   epr performs regularized, iterative image restoration with a non-
   negativity constraint.  The before-image is a three dimensional (3-D) 
   TIFF image composed of rectangular, regularly spaced, optical sections.
   Large images are decomposed into smaller, overlapping (in x and y only) 
   image segments for restoration, and the restored segments are recomposed.
   The after-image is the restored 3-D image. All options must appear 
   before the image file names, but the order of options is not important.

Example Usage

epr -psf mypsf_ -smoothness 0.0005 -iterations 250 mycell_ mycell_r